37. SPECIFICATIONS
298
Pressure input range: 500 to 1,400 hPa (in 1hPa step)
375 to 1,050 mmHg (in 1mmHg step)
14.8 to 41.3 inchHg (in 0.1inchHg step)
ppm input range: -499 to 499 ppm (in 1 ppm step)
Prism constant correction: -99 to 99 mm (in 1 mm step)
0mm fixed for reflectorless measurement
Earth curvature and refraction correction:
No/Yes K=0.142/Yes K=0.20 (selectable)
Scale factor setting: 0.5 to 2.0
Sea level correction: No/Yes (selectable)
*1: Slight haze, visibility about 20 km, sunny periods, weak scintillation.
*2: No haze, visibility about 40 km, overcast, no scintillation.
*3: Measurement at -30 to -20°C (-22 to -4°F) (Low Temperature Model)/50
to 60°C (122 to 140°F) (Standard Model)
*4: The temperature of the lower limit is -30°C ( -22°F) when using reflective
sheet target. (Low Temperature Model)
*5: Figures when measured at -30°C (-22°F) or more. (Low Temperature
Model)
*6: Figures when the laser beam strikes within 30° of the reflective sheet
target.
*7: Figures when using Kodak Gray Card White side (reflection factor 90%)
and brightness level is less than 30000 lx (a little cloudy).
*8: Figures when using Kodak Gray Card Gray side (reflection factor 18%)
and brightness level is less than 30000 lx (a little cloudy).
*7,*8: When performing reflectorless measurement, the possible measurement
range and precision will change depending on the target reflection factor,
weather conditions and location conditions.
*9: Face the prism toward the instrument during the measurement with the
distance at 10 m or less.
*10: At the time of the tracking measurement.
Guide Light
Light source: LED (red 626 nm/green 524 nm)
Distance: 1.3 to 150m
*1
Visible range: Right and Left/Upward and Downward:
± 4° (7m/100m)
Resolving power at center area (width):
4’ (about 0.12m/100m)
Brightness 3 levels (bright/normal/dim)
Internal memory
Capacity 10,000 measurement points