EasyManua.ls Logo

Topcon SR-LEDW - Page 107

Topcon SR-LEDW
140 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
- 96 -
(3) “STB” Command
When the measurement is conducted by order of STB command, the measured data is returned
with binary data (big endia) from the instrument.
Measurement Output Format
‘4.2.1 Remote Measurement Output Format’
· Measured data of spectral radiance is returned at all time for STB command even if “D1”
command is transmitted.
1
Header
Header have necessary information to receive data.
Initial bite
Function
Data type
Size
Qty of
element
Remark
1
Size of data
Unsigned
integer
4
1
5
Checksome data
Unsigned
integer
1
1
Data size
5B(Bite)
Checksome dataThe value from which 1 byte of lower rank of the total that has added data in turm
from first by the byte unit was picked out.
2
Data (When finish measurement
Initial bite
Function
Data type
Size
Qty of
element
Remarks
1
Measurement angle
Unsigned
integer
1
1
1:2°,2:1°,3:0.2°,4:0.1°
2
Integral time
Floating point
4
1
6
Radiance
Floating point
4
1
10
Luminance
Floating point
4
1
14
Tristimulus values X
Floating point
4
1
18
Tristimulus values Y
Floating point
4
1
22
Tristimulus values Z
Floating point
4
1
26
Chromaticity x
Floating point
4
1
30
Chromaticity y
Floating point
4
1
34
Chromaticity u’
Floating point
4
1
38
Chromaticity v’
Floating point
4
1
42
Correlated Color
temperature Tc
Floating point
4
1
If TC can not be
calculated, instrument
send a value of -1
46
Deviation duv
Floating point
4
1
If TC can not calculated,
instrument send a value
of -1.
50
Wave length
Unsigned
integer
2
1
380nm (fix)
52
Spectral Radiance
Floating point
4
1
Spectral Radiance at
380nm
56
Wave length
Unsigned
integer
2
1
381nm (fix)
58
Spectral Radiance
Floating point
4
1
Spectral Radiance at
381nm
Note

Table of Contents

Related product manuals