Chapter 12: Troubleshooting Issues 00.053.129, Revision C SonixTablet Service Manual
12-6
12.13.5 System Hangs when Loading Maps (at Initialization)
The initialization process hangs when loading Maps:
A.7.3 System Hangs When Loading Maps (at Initialization).
12.13.6 Successful Initialization but only 4DC7-3/40 Transducer is Detected
If the system passes all error checks during initialization, but no transducers are detected except the 4DC7-3/40 (if
it is connected), contact Ultrasonix Technical Support.
12.14 NO NETWORK CONNECTION VIA ETHERNET CABLE
If the system is connected to the network via an Ethernet cable but a network connection cannot be established, there may
be an issue with the network cable connection or with the BIOS.
Check the BIOS Network configuration as the factory BIOS network setting is insufficient when the modulo has two (2)
network ports:
A.8.1 BIOS Network Configuration (for two (2) Network Ports).
12.15 TRANSDUCER CONNECTION PORT NOT FUNCTIONING
If a transducer connection port is not functioning, the MUX board may need to be replaced:
A.9 Transducer Connection Port not Functioning.