Chapter 3: System Installation 00.053.129, Revision C SonixTablet Service Manual
3-8
3.3 INSTALLATION
3.3.1 EMI Filter and Power Pack
3.3.1.1 Voltage Configuration
The diagram below shows the location of the EMI Filter which houses the system's fuses. Refer to 3.3.2.2 for
details on how to remove the fuse box and replace the fuse(s).
When changing the voltage settings, the EMI filter must be adjusted to the relevant setting and the fuses must be
flipped to match the selected voltage.
Figure 3-2: EMI Filter
Figure 3-3: EMI Filter and System Power Button Location
Caution: The system voltage setting is configured in the factory. It should not need to be changed in the field.
Main Power Switch
Power Cord (to Wall Outlet) Receptacle