EasyManua.ls Logo

Ultrasonix SonixTablet - Cleaning;Disinfecting Non-Invasive Transducers

Ultrasonix SonixTablet
123 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 11: Maintenance 00.053.129, Revision C SonixTablet Service Manual
11-10
11.3.5 Cleaning/Disinfecting Non-Invasive Transducers
To prevent biological materials (bioburden) from drying on the equipment, always reprocess transducers as soon
as is reasonably practical following use.
Table 11-3: Non-Invasive Cleaning/Disinfecting Agents
NON-INVASIVE
T
RANSDUCERS
CLEANING/DISINFECTING AGENTS
75% IPA
Cidex Activated Dialdehyde
Solution 14 day
Cidex Plus 28 day
Cidex OPA
Cidezyme
Klenzyme
McKesson Brand
Metrizyme
Milton Disinfecting Liquid
Nuclean
Omnicide - FG2
T-spray
MISCELLANEOUS
SA4-2/24
PA7-4/12
♦♦
MC9-5/12
HST15-8/20
♦♦♦♦♦
4DC7-3/40
PA4-2/20
♦♦♦♦♦
GPS
C5-2/60 GPS
L14-5/38 GPS
♦♦♦ ♦♦♦ ♦♦♦
LINEAR
AND CONVEX
C5-2/60
C7-3/50
♦♦♦♦♦♦♦ ♦♦
L9-4/38
L14-5/38
♦♦♦♦♦♦♦ ♦♦
L14-5W/60
Caution: Use only Ultrasonix recommended cleaners/disinfectants (Table 11-3). They have been tested and
determined safe to use on Ultrasonix transducers. Failure to follow these instructions may cause damage
and will void transducer warranties.

Table of Contents

Related product manuals