Appendix A: Troubleshooting Solutions 00.053.129, Revision C SonixTablet Service Manual
A-2
A.3.3 Check DICOM Settings
To Check the DICOM Settings:
1. Tap the touch screen button.
1. Tap the touch screen Menu button.
2. Select Administrator > DICOM.
3. Check to ensure that none of the Turn on DICOM… settings are selected.
4. Select OK.
A.3.4 Clear DICOM Settings
To Clear the DICOM Queues:
1. Tap the touch screen button.
2. Select Administrator > Service Mode.
3. Enter the Service Password.
4. In the Service dialog beside Windows Explorer, select Go.
5. Go to D:\PateintInfo.
6. Look for file dcm_storage_queue.xml and dcm_print_queue.xml.
7. Right click on the two files using the Update key then select Delete.
8. Reboot the machine.
A.3.5 Delete Older LUTS Files
1. Ensure a USB keyboard is connected to the system.
2. Press the WINDOWS START KEY and select All Programs > Accessories > Windows Explorer.
3. Navigate to D:\Ultrasonix Settings\LUTS.
4. Delete the LUTS file.
A.3.6 Delete Older LOG Files
1. Ensure a USB keyboard is connected to the system.
2. Press the WINDOWS START KEY and select All Programs > Accessories > Windows Explorer.
3. Navigate to D:\Ultrasonix Settings\LUTS.
4. Delete the Log file.
Note: This option applies only to software versions 5.4.3 and earlier.