SonixTablet Service Manual 00.053.129, Revision C Chapter 7: Field Service Components
7-1
CHAPTER 7: FIELD SERVICE COMPONENTS
This section describes how to handle all field serviceable components, including the modulo, LCD display, console, speakers, UPS,
battery and transducers.
7.1 PROTECTING PATIENT DATA
Before doing any work with the modulo, be sure to complete sections 7.1.1 and 7.1.3 (or 7.1.5) in order to the export/backup
and therefore protect the integrity of each client's data.
Once all work has been completed, refer to sections 7.1.2 and 7.1.4 (or 7.1.6) to import/restore the client's settings and data.
7.1.1 Exporting User Data (As Required)
Before servicing the modulo, export User Data to an external USB backup medium (memory stick, external hard
drive, etc.). This will ensure access to this data if either the modulo or hard drive needs replaced during servicing.
The following options are available for export:
• Imaging Presets
• 3D/4D Presets
• DICOM Server Configuration
• Settings
• Measurement Order
• Obstetrical Tables
•Protocols
• General
• Preset Assignments
• Measurement Customization
• Worksheets
•Banners
• System Logs
• Licenses (license.txt file).
Note: Use only one (1) method to Export/Import Patient/Exam data:
• when exporting with 7.1.3, import with 7.1.4
• when backing up with 7.1.5, restore with 7.1.6.
Note: When creating a backup prior to servicing or replacing the modulo, Ultrasonix recommends selecting all
options for Export.
Note: General is the only Protocol available on SonixTablet systems.
Note: System Logs cannot be imported.