SonixTablet Service Manual 00.053.129, Revision C Chapter 7: Field Service Components
7-9
7.1.6 Restoring/Importing Patient Data via the Service Mode Option
In order to restore Patient/Exam data via Service Mode, it must first have been backed up using this option. Refer
to 7.1.5 for backup details.
During the restoration process, the system will warn – in a multi-step process – about overwriting any patient data
that already exists on the system hard drive.
Equipment/Tools Required:
USB keyboard
USB media (memory stick, external hard drive, etc.) containing previously backed up Patient/Exam data.
To Restore/Import Patient Data (via the Service Mode Option):
1. Connect the USB keyboard to the one of the four (4) USB ports.
2. Connect the USB storage device to the one of the three (3) remaining USB ports.
3. Tap the touch screen button.
4. Select Administrator > Service….
5. Type in the Service Mode Password using the USB keyboard.
6. Select OK.
7. From the Advanced Tools and Options drop-down menu, select Patient Data Backup/Restore.
8. Select Go.
Cautions:
The Restore function will delete any patient data that exists on the system.
Due to privacy regulations and laws, Service Technicians are forbidden from removing patient data from any client
site. Refer to 1.4 Privacy Disclaimer and 7.2.1 Removing the Hard Drives for more details.
Note: USB storage space required to backup patient data depends on the number of patients/images on the
system. Hard drive capacity is 160 Gb in total.
Note: The system will auto-detect the USB storage device.