Page 122
Programming a Flow Test
(Continued)
9. Press the NEXT keypad to highlight Max Flow. Maximum flow is
the high setpoint Sprint LC uses to decide whether the test was a
pass or fail. Use the INCREMENT and DECREMENT keypads to set the
value for maximum flow.
10. Press the NEXT keypad to highlight Min Flow. Minimum flow is the
low setpoint Sprint LC uses to decide whether the test was a pass
or fail. Use the INCREMENT and DECREMENT keypads to set the
value for minimum flow. Press BACK to exit back to the main
parameter menu. The minimum flow cannot be set higher than the
maximum flow setpoint.
11. Press the NEXT keypad to highlight Next Program. Use the
INCREMENT and DECREMENT keypads to set the number of the next
program to link. If not linking to another program, set the next
program to zero.
12. Press the BACK keypad after all settings are entered and ready.
After pressing the BACK keypad, Sprint LC returns to the setup
menu.