Calibration Access
Procedure
1. Turn the LOCKOUT keyswitch to program position. The startup menu
is displayed.
2. Press the NEXT keypad to highlight the calibration option.
3. Press the ENTER/BACK keypad to select calibration. A flashing warning
message will appear as shown in Figure 75.
Flashing Access Warning
The warning encountered when reaching the first calibration selection
menu flashes until the operator takes one of two possible actions—
Stop and Leave Calibration
To quit proceeding with calibration, exit by pressing the ENTER/BACK
keypad when the Exit Calibration line is highlighted. Sprint LC defaults to
the Exit Calibration line.
Proceed with Calibration
To proceed with calibration, use the NEXT keypad to select the Enter
Calibration line, then press the ENTER/BACK keypad. After the ENTER/BACK
keypad is pressed, Sprint LC displays the menu in Figure 76.
Verify First
If calibration is within tolerances there is no
need re-set the calibration. The calibration
tolerances can be found on the calibration
report that was shipped with the Sprint LC.