Page 56
Pneumatic Circuit
The way the Sprint LC is programmed determines which valves are
activated and in what order valves open and close. All, or merely a few,
components in this circuit might be used in a testing application.
FIGURE 22
Self-Contained
Sprint LC has all valves and pneumatic components in one package to
perform pressure decay, occlusion, flow, and burst testing, if so equipped.
Pressure
Regulator
Flow
Sensor
Flow-A
Valve #1
Press.
X-ducer
TestPort
(Port 1)
Flow
Control
Flow-B
Valve #5
Pressure Decay
Valve #3
Couple-A
Valve #8
Coupling Port
(Port 2, or on rear of tester)
Product
Test
Fixture
Burst-A
Valve #4
Burst-B
Valve #6
Dump
Valve #2
Air
Atmosphere