Pressure Calibration Example
FIGURE 78
Target value on Sprint LC Readout on external
calibration menu pressure standard
0.00 0.000
5.00 5.000
10.00 10.000
15.00 15.000
Adjust the pressure regulator on Sprint LC to make the pressure standard
match the target value on the calibration menu.
Backward Calibration
A common mistake is to match the pressure
(accept the calibration pressure) by watching the
“reading” value on Sprint LC. Instead, match the
pressure from the pressure standard to the target
number offered on the Sprint LC.
Match
!
U
s
o
n
S
PR
I
N
T
L
C
1
2
S
p
rint LC
Pressure
Standard