Page 84
Explanation of Test Result Messages
PASS
The product stayed within all test parameters of
each test that was run.
ABORT The START switch was pressed, ending the test.
P ERR (Pressure Error) The test pressure was above or
below the pressure error window.
GROSS The test pressure dropped below the pressure error
limit during the stabilize time.
LEAK The pressure decay exceeded the reject level during
test time in a pressure decay test.
OCC (Occlusion) In an occlusion test, the pressure decay
did not reach the reject level setpoint.
HIGH The product exceeded the maximum flow or burst
setpoint.
LOW The product did not meet the minimum flow or
burst setpoint.
OVER (Over Pressure) In a flow test, the pressure was
above the maximum pressure the flow sensor can
withstand.
NO (No Burst) The burst test time elapsed before a
burst event was detected.