In Case of Trouble
If there is difficulty setting up the Sprint LC, please read this guide
completely. Many “problems” can be solved with a few minor adjustments
to the tester. Try the items on this checklist first. If all else fails, call Uson
for help.
Setup Checklist
Is supply air connected to Sprint LC?
Is supply air clean and dry? Air that contains moisture, oil, or hard
particulate matter WILL damage Sprint LC.
Is the supply air within the correct pressure range?
Was the test port capped while setting test pressure? This must be
done to get the correct setting.
Has the pressure regulator changed from when the test pressure
was set? A change will cause pressure errors.
Is the display contrast correctly set using the control on the back?
Is the tone volume correctly set using the control on the back?
Is the LOCKOUT keyswitch in the correct position?
Is the product attached to the test port during all flow tests?
Without product, the resulting high flow could damage the flow
sensor.
Has the calibration been changed? Do not enter calibration mode
unless intending to reset calibration data and the necessary
standards and instruments to complete the task are at hand.
Are engineering units set in the correct units of measure?