STAB: Abbreviation for stabilization. A time interval following fill phase
that allows product to settle before Sprint LC starts the measurement
phase. Longer stabilization times are often required in products
constructed of flexible materials.
Page 59, 97
Start Switch: The square white pushbutton directly under the DECREMENT
keypad. The START switch has two functions. Function one: Initiates the
test. Function two: Press the START switch during a test to abort the
current test.
Page 8-9
Startup Display: The first display that briefly appears on Sprint LC’s
LCD when the power is switched on. This display shows the available test
modes, the serial number and the version of firmware running the tester.
Page 28
Stored Programs: A set of instructions (parameters) that can be set by the
customer to run a variety of tests. Sprint LC is shipped with four stored
(factory-installed) programs. There are a total of 99 programs available.
Users can alter stored programs to meet specific product testing needs.
Programs are configured in Sprint LC’s parameters menu and are kept in
non-volatile RAM.
Page 34
Stored Programs, changing from one to the other: In run mode, press
and hold the PREVIOUS or NEXT keypads to change the current program to
another stored program.
Page 32
Supply Air: Compressed air or gas connected to the back fitting labeled
Air Supply. Sprint LC’s standard fitting is a 1/8 inch female NPT
bulkhead. Air must be clean, dry, and free of oil. Unless the tester is a
custom, the maximum supply pressure must not exceed 150 psig.
Page 22-23
Page R14