EasyManuals Logo

YOKOGAWA AQ7280 User Manual

YOKOGAWA AQ7280
195 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #162 background imageLoading...
Page #162 background image
7-2
IM AQ7280-01EN
7.2 Initializing the AQ7280 to Its Factory Default
Condition
Procedure
Initialization Screen
1.
Press SETUP and then the System Setup soft key.
2.
Using the rotary knob and ENTER, select Factory Setting to display the following screen.
Executes setup initialization
Measurement, analysis, OTDR, and system settings are
initialized to their factory default values.
Executes all initialization
All settings are initialized to their factory default values.
The internal memory will also be formatted.
Explanation
You can initialize the AQ7280 settings to their factory default conditions.
Executing Setup Initialization
Measurement, analysis, OTDR, and system settings are initialized to their factory default values. The
internal memory and the day and time settings (excluding the display type) are not initialized.
Executing All Initialization
All settings are initialized to their factory default values. The internal memory will also be formatted,
and all data will be erased. If necessary, back up the data before initializing. The day and time settings
(excluding the type) are not initialized.
Note
If you send files and folders in the AQ7280 internal memory to a PC using the AQ7280 mini-B USB port, they
are downloaded at the data rate depending on the PC performance. To use this feature, the AQ7280 mini-B
USB port function must be set to Storage (see section 7.6).

Table of Contents

Other manuals for YOKOGAWA AQ7280

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA AQ7280 and is the answer not in the manual?

YOKOGAWA AQ7280 Specifications

General IconGeneral
BrandYOKOGAWA
ModelAQ7280
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals