EasyManua.ls Logo

YOKOGAWA AQ7280 - Measuring Splice Loss; When Marker Mode Is Set to Line; When Marker Mode Is Set to Marker

YOKOGAWA AQ7280
195 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4-9
IM AQ7280-01EN
Optical Pulse Analysis
1
2
3
4
5
6
7
8
Index
Measuring Splice Loss
When Marker Mode Is Set to Marker
1.
From the top menu, select OTDR.
2.
Press the Marker soft key and then the 4 Point Markers soft key to display the following
screen.
Measurement start position
Set marker .
1
2
Splice loss event start position
Set marker .
Splice loss event end position
Set marker Y2.
Measurement end position
Set marker .
3
Measured splice loss (result)
When Marker Mode Is Set to Line
1.
From the top menu, select OTDR.
2.
Press the Marker soft key and then the 5 Point Markers soft key to display the following
screen.
Near-end point
Set marker n.
Approximated line point on
the near-end side
Set marker N.
Splice loss detection point
Set marker E.
Approximated line point on
the far-end side
Set marker F.
Far-end point
Set marker f.
Measured splice loss (result)
4.1 Analyzing Waveforms

Table of Contents

Other manuals for YOKOGAWA AQ7280

Related product manuals