7-1
IM WT18
O
1E-02EN
Chapter 7 Computations
7.1 Setting User-Defined Functions
This section explains the following settings for user-defined functions:
Turning computations on and off
Computation name
• Unit
Expressions
Turning max hold on and off
“User-Defined Functions (User Defined Function)” in the features guide
Setting User-Defined Functions (User Defined Function)
Press MEASURE and then the User Defined Function soft key to display the following screen.
Displays the setup screen for
user-defined functions F1 to F5
Displays the setup screen for
user-defined functions F6 to F10
Displays the setup screen for
user-defined functions F11 to F15
Displays the setup screen for
user-defined functions F16 to F20
Turns max hold on and off
Set the computation
name
(up to 8 characters).
Turns the computation
on and off
Set the expression.
Set the unit
(up to 8 characters).