Page 82 February 2020TD 266 OPERATING MANUAL ZPulse® DCS 4420/4830/4520/4930
User Maintenance is a password protected menu that contain settings
that normally are not altered by non trained personel. However if you
need to change some settings here the password will be 1000 for all units
Enter the password and press OK og select the read only to only check
the settings without risk of incidentaly change one of the properties.
Figure 5-15: User Maintenance Password
User Maintenance are devided into 7 sections
Sheet 1/7 contains Mandatory settings.
All sensors are given a Node Description text like DCS #xxx (xxx is the
serial number of the sensor). The user can modify this node description
text if required. Be aware that the node description changes to *Corrupt
Configuration if it has lost the configuration in flash. Contact the factory if
this happens. The configuration is saved in two sectors in flash memory.
A flash sector can be corrupted if the power is lost during the saving of
new configuration. The double flash sector saving ensures that it does not
lose the configuration. If one of the sectors is corrupted, the other sector
is used and also saved to the corrupt sector.
Figure 5-16: Mandatory