EasyManuals Logo

Agilent Technologies 6890 Series Service Manual

Agilent Technologies 6890 Series
867 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #224 background imageLoading...
Page #224 background image
17 of 24
Jun 2001
Detectors
Agilent 6890 Gas Chromatograph Service Manual
Flame Ionization Detector (FID) 310
Replacement procedures
Figure 310-13 Bracket mounted onto the manifold
3. If the tubing from the gang fitting bends directly to the left, re-shape it
so that it bends up and away from the block as shown in Figure 310-15.
4. Insert the gang fitting through the cutout in the manifold bracket and
install it onto the new manifold assembly so that the tubing runs back
and away from the fitting. Be sure the left tube clears the inner edge of
the bracket. See Figure 310-13 and Figure 310-15.
Finger-tighten nuts.
Do not use a wrench.
(FPD shown above)
7/16-inch hex nuts
Mounting bracket
Gang fitting
Label tag through
Ribbon cable
installs here
slot in bracket

Table of Contents

Other manuals for Agilent Technologies 6890 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 6890 Series and is the answer not in the manual?

Agilent Technologies 6890 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model6890 Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals