Agilent 6890 Gas Chromatograph Service Manual
Auxiliary EPC 360
Replacement procedures
3. Pull the frit to be changed out of the block. Also remove the O-ring that
seals it.
4. Place an O-ring on the new frit. Place the O-ring/frit combination in the
block.
5. Reconnect the block to the pneumatics module. Tighten the screw firmly.