830 Warning Codes
Warning 109Front inlet: config changed, method defaulted
4 of 10
Jun 2001
Error Codes and Internal Diagnostics
Agilent 6890 Gas Chromatograph Service Manual
Warning 109Front inlet: config changed, method defaulted
Warning 110Back inlet: config changed, method defaulted
Status message: F inl config changed or B inl config changed
This message appears when:
The 6890 GC is turned on after a new inlet has been installed.
The 6890 GC is turned on and senses that the inlet has a different type
of sensor than was there when the instrument was turned off.
A method is loaded that calls for a different type of inlet than the one
currently installed.
The following will occur:
If a new inlet or sensor was installed, the 6890 GC will enter the default
setpoints for the new type of inlet.
If a method was loaded with a different inlet specified, the 6890 GC will
not change the method but will flash the warning message.
Warning 111Column 1: config changed, method defaulted
Warning 112Column 2: config changed, method defaulted
Status message: Col 1 config changed or Col 2 config changed
This message appears when:
The 6890 GC is turned on after a different column has been installed.
You load a method that calls for a different type of column than the one
currently installed in the oven.
The following will occur:
If a new column was installed, the 6890 GC will enter the column default
setpoints.
If a method was loaded for a different type of column than the one
installed, the 6890 GC will not change the method but will flash the
warning message.