Error Codes and Internal Diagnostics
Agilent 6890 Gas Chromatograph Service Manual
Faults 850
Fault 205Signal DSP RAM broken
Fault 205Signal DSP RAM broken
Status message: Sig DSP RAM broken
The detectors signal processing electronics are not functioning correctly.
This indicates a malfunction with the 6890 GC main PC board. The signal
path will not function.
Turn the instrument on and then off at least one time. If the error still
occurs, the mainboard must be replaced.
Fault 206Signal DSP registers
Status message: Sig DSP registers
The detectors signal processing electronics are not functioning correctly.
This indicates a malfunction with the 6890 GC main PC board. The signal
path will not function.
Turn the instrument on and then off at least one time. If the error still
occurs, the mainboard must be replaced.
Fault 207Sig DSP data corrupt
Status message: Sig DSP data corrupt
The detectors signal processing electronics are not functioning correctly.
This indicates a malfunction with the 6890 GC main PC board.
Turn the instrument on and then off at least one time. If the error still
occurs, the mainboard must be replaced.