Error Codes and Internal Diagnostics
Agilent 6890 Gas Chromatograph Service Manual
Faults 850
Fault 279Pneumatics aux: invalid constants from factory
Fault 279Pneumatics aux: invalid constants from factory
Status message: P aux bad fact cal
The aux pneumatics modules calibration is invalid.
Fault 280F inlet read/write failure. Module is unusable.
Fault 281B inlet read/write failure. Module is unusable.
Status message: F inlet i/o failure or B inlet i/o failure
The EPC module is malfunctioning. Replace the module.
Fault 282F det read/write failure. Module is unusable.
Fault 283B det read/write failure. Module is unusable.
Status message: F det i/o failure or B det i/o failure
The EPC module is malfunctioning. Replace the module.
Fault 284Pneu aux read/write failure. Module is unusable.
Status message: Pneu aux i/o failure
The EPC module is malfunctioning. Replace the module.
Fault 285Front detector offset adjustment failed
Fault 286Back detector offset adjustment failed
Status message: F det adjust failure or B det adjust failure
An ECD or NPD detector has failed to successfully achieve the target offset
setpoint within the adjustment range. The instrument will not be ready until
Adjust Offset is turned Off or On.
Fault 287F OIM not installed
Fault 288B OIM not installed
The OIM module is required by the method but is not present.