EasyManuals Logo
Home>Agilent Technologies>Laboratory Equipment>7100

Agilent Technologies 7100 User Manual

Agilent Technologies 7100
266 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #131 background imageLoading...
Page #131 background image
Agilent 7100 Capillary Electrophoresis System User Manual 131
Using the Agilent 7100 Capillary Electrophoresis System 5
Capillary Electrophoresis, CE
Closing the Capillary Cassette
1 Place the top of the cassette onto the bottom part with an offset of 8 mm
(0.3 inch).
2 Hold the bottom part and slide the top part towards the top of the cassette
until it clicks into place.
3 Make sure the cassette cover and body form one unit
4 Make sure the two capillary ends outside the cassette are both the same
length as the cassette guiding pins and are parallel to them. Use the
capillary lock to make adjustments to the inlet side of the capillary. The
capillary end on the outlet side is fixed; it cannot be moved.

Table of Contents

Other manuals for Agilent Technologies 7100

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7100 and is the answer not in the manual?

Agilent Technologies 7100 Specifications

General IconGeneral
BrandAgilent Technologies
Model7100
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals