Agilent 7100 Capillary Electrophoresis System User Manual 131
Using the Agilent 7100 Capillary Electrophoresis System 5
Capillary Electrophoresis, CE
Closing the Capillary Cassette
1 Place the top of the cassette onto the bottom part with an offset of 8 mm
(0.3 inch).
2 Hold the bottom part and slide the top part towards the top of the cassette
until it clicks into place.
3 Make sure the cassette cover and body form one unit
4 Make sure the two capillary ends outside the cassette are both the same
length as the cassette guiding pins and are parallel to them. Use the
capillary lock to make adjustments to the inlet side of the capillary. The
capillary end on the outlet side is fixed; it cannot be moved.