Agilent 7100 Capillary Electrophoresis System User Manual 157
Using the Agilent 7100 Capillary Electrophoresis System 5
Capillary Electrochromatography, CEC
Timetable Options with CEC
The installed high pressure supply may also be used in the general timetable of
a method setup. The Pressure command is extended by the External pressure
functionality, see Figure 49 on page 157.
Figure 49 Pressure options in CEC mode
Timetable-controlled changes in the pressure settings are:
• OFF: Pressure usage is switched off.
• Internal, pressure (-100 mbar to +100 mbar): Can be applied to the inlet
vial.
• Flush: Standard flush procedure is applied with an internally generated
pressure of approximately 950 mbar.
• External: Allows pressurizing the inlet vial and/or the outlet vial. External
pressure is reduced to the set value between 2 and 12 bar and applied in a
controlled manner.