Agilent 7100 Capillary Electrophoresis System User Manual 253
Appendix 10
Setting Up a Test Sample Run
Method Parameter CE (Default Settings)
Vials
Inlet home vial 4
Outlet home vial 5
Cassette Temperature
20.0°C
High Voltage System
Enable high voltage
Voltage 0.0 V (will ramp up in accordance with timetable)
Current 300 µA
Power 6.0 W
Low current limit OFF
Stoptime
10.0 minutes
Posttime
OFF
Replenishment
Replenishment and preconditioning parallel processing
No replenishment used