Agilent 7100 Capillary Electrophoresis System User Manual 105
Using the Agilent 7100 Capillary Electrophoresis System 5
Capillary Electrophoresis, CE
Using an Injection Program
The injection table is used for advanced injection tasks such as:
Injection from different vials, for example ITP.
Multiple injection modes, for example spiking.
Injecting a buffer plug after the sample to prevent sample loss after
applying voltage.
The online help provides more information on this.
Timetable
In the CE Timetable, present on the right-hand side of the CE Method Setup dialog
you can choose to set and change values for the following parameters at a
certain time during the run:
Voltage
Current
Power
Pressure
Inlet vial
Outlet vial
Cassette temperature
Lower alarm limit for current
High pressure (only in CEC mode)
Time programming of these parameters can be useful for example to:
Define gradients (for voltage or temperature).
Change the buffer during the analysis (inlet and outlet).
Apply pressure during the analysis.
Change the outlet vial during analysis.
Change the alarm limit for current for example, when changing the buffer
system during the analysis.