70 Agilent 7100 Capillary Electrophoresis System User Manual
4 LAN Configuration
Storing the Settings Permanently with Bootp
Storing the Settings Permanently with Bootp
If you want to change parameters of the CE instrument using Bootp follow the
instructions below.
1 Turn off the CE instrument.
2 Change the CE instrument's settings for the configuration switch to “Bootp
& Store” mode, see Table 18 on page 65.
3 Start the Agilent Bootp Service (see “Agilent Bootp Service” on page 75)
and open its window.
4 If required, modify the parameters for the CE instrument according to your
needs using the existing configuration.
5 Press OK to exit the Bootp manager.
6 Now turn on the CE instrument and view the Bootp server window. After
some time the Agilent Bootp Service will display the request from the LAN
interface. The parameters are now stored permanently in the non-volatile
memory of the CE instrument.
7 Close the Agilent Bootp Service and turn off the CE instrument.
8 Change the settings of the CE instrument’s configuration switch to “Using
Stored” mode, see Table 18 on page 65.
9 Power cycle the CE instrument. The CE instrument can be accessed now via
LAN without the Agilent Bootp Service, refer to “Agilent Bootp Service” on
page 75.

Table of Contents

Other manuals for Agilent Technologies 7100

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7100 and is the answer not in the manual?

Agilent Technologies 7100 Specifications

General IconGeneral
BrandAgilent Technologies
Model7100
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals