216 Agilent 7100 Capillary Electrophoresis System User Manual
8Maintenance
Maintenance of the Replenishment System
Maintenance of the Replenishment System
The replenishment system automatically empties and refills the buffer vials in
the tray with new buffer from the electrolyte bottle. Typically, replenishment
is done before every run or after every 3 to 4 runs, depending on the stability
and volume of the running buffer. Initial preparation of the replenishment
system is described in “Preparing the Replenishment System if Needed” on
page 111. Please see the general instructions on operations for replenishment
maintenance such as flush tubes, clean tubes and clean level sensor
(“Operations for Replenishment Maintenance” on page 219).
Frequency
The replenishment system (bottles and tubing) should be cleaned if:
using it for the first time,
changing the buffer, or
the replenishment system will be idle for some time.
Requirements
Vials and caps
Water and/or buffer and isopropanol (depending on the protocol selected)
ChemStation, Lab Advisor or Instrument Utilities software

Table of Contents

Other manuals for Agilent Technologies 7100

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7100 and is the answer not in the manual?

Agilent Technologies 7100 Specifications

General IconGeneral
BrandAgilent Technologies
Model7100
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals