Agilent 7100 Capillary Electrophoresis System User Manual 63
LAN Configuration 4
TCP/IP Parameter Configuration
TCP/IP Parameter Configuration
To operate properly in a network environment, the LAN interface must be
configured with valid TCP/IP network parameters. These parameters are:
IP address
Subnet mask
Default gateway
The TCP/IP parameters can be configured by the following methods:
by automatically requesting the parameters from a network-based BOOTP
service (using the so-called bootstrap protocol), see “Agilent Bootp
Service” on page 75
by manually setting the parameters using Telnet
The LAN interface differentiates between several initialization modes. The
initialization mode (abbreviation: 'init mode') defines how to determine the
active TCP/IP parameters after power-on. The parameters may be derived
from a Bootp cycle, non-volatile memory or initialized with known default
values. The initialization mode is selected by the configuration switch, see
Table 17 on page 64.

Table of Contents

Other manuals for Agilent Technologies 7100

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7100 and is the answer not in the manual?

Agilent Technologies 7100 Specifications

General IconGeneral
BrandAgilent Technologies
Model7100
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals