Agilent 7100 Capillary Electrophoresis System User Manual 71
LAN Configuration 4
Manual Configuration
Manual Configuration
Manual configuration only alters the set of parameters stored in the
non-volatile memory of the CE instrument. It never affects the currently active
parameters. Therefore, manual configuration can be done at any time. A
power cycle is mandatory to make the stored parameters become the active
parameters, if the initialization mode selection switches permit this.
Figure 7 Manual configuration (principle)