Agilent 7100 Capillary Electrophoresis System User Manual 65
LAN Configuration 4
Initialization Mode Selection
Initialization Mode Selection
The following initialization (init) modes are selectable:
Bootp
When the initialization mode "Bootp" is selected, the 7100 CE tries to
download the parameters from a Bootp server. The parameters obtained
become the active parameters immediately. They are not stored in the
non-volatile memory of the 7100 CE. Therefore, the parameters are lost with
the next power cycle of the 7100 CE.
Figure 3 Bootp (principle)
Table 18 Initialization mode switches
SW6 SW7 SW8 Init mode
OFF OFF OFF Bootp
OFF OFF ON Bootp & Store
OFF ON OFF Using Stored
OFF ON ON Using Default