EasyManua.ls Logo

Agilent Technologies 7100 - Page 66

Agilent Technologies 7100
266 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
66 Agilent 7100 Capillary Electrophoresis System User Manual
4 LAN Configuration
Initialization Mode Selection
Bootp & Store
When "Bootp & Store" is selected, the parameters obtained from a Bootp
server become the active parameters immediately. In addition, they are stored
in the non-volatile memory of the 7100 CE. Thus, after a power cycle they are
still available. This enables a kind of "Bootp once" configuration of the 7100
CE.
Example: The user may not want to have a Bootp server be active in his
network all the time. But on the other hand, he may not have any other
configuration method than Bootp. In this case he starts the Bootp server
temporarily, powers on the 7100 CE using the initialization mode "Bootp &
Store", waits for the Bootp cycle to be completed, closes the Bootp server and
powers down the 7100 CE. Then, he selects the initialization mode "Using
Stored" and powers on the 7100 CE again. From now on, he is able to establish
the TCP/IP connection to the 7100 CE with the parameters obtained in that
single Bootp cycle.
Figure 4 Bootp & Store (principle)
NOTE
Use the initialization mode "Bootp & Store" carefully, because writing to the non-volatile
memory takes time. Therefore, if the 7100 CE is to obtain its parameters from a Bootp
server every time it is powered on, the recommended initialization mode is "Bootp"!

Table of Contents

Other manuals for Agilent Technologies 7100

Related product manuals