Agilent 7100 Capillary Electrophoresis System User Manual 225
Maintenance 8
Changing the Lamp
Removing the Existing Lamp
1 Select Exit from the File menu to exit the Agilent ChemStation. Exit both
online and offline copies.
2 Turn off the line power to the Agilent CE instrument.
3 Open the top cover.
4 Open the detector cover.
Parts required Pozidriv screwdriver (part of the accessory kit).
Deuterium lamp for the Agilent CE instrument.