EasyManuals Logo
Home>Agilent Technologies>Laboratory Equipment>7100

Agilent Technologies 7100 User Manual

Agilent Technologies 7100
266 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #166 background imageLoading...
Page #166 background image
166 Agilent 7100 Capillary Electrophoresis System User Manual
6 Troubleshooting and Diagnostics
CE Problem Solving
Broken Capillary
To check for a broken capillary, remove the capillary cassette and open it as
described under “Removing the Capillary Cassette from the Agilent CE
Instrument” on page 121. The capillary is especially sensitive to breaking at
the detection window and the capillary ends, therefore check these parts of
the capillary. If the capillary is broken at the detection window, a leakage
current of nearly the same value as the expected current may occur.
Small instabilities in the current curves, as shown in Figure 51, may also be a
sign of a broken capillary.
Figure 51 Current instabilities
If the capillary is broken at the ends, it may no longer reach into the liquid in
the vial. In that case, the analysis is stopped if the low current alarm limit has
been reached (this must be set first). Sometimes the capillary has very small
cracks which are hardly recognizable under a microscope.
Exchange the broken capillary
Contaminated Insulation Plate
Salt crystals or other contamination of the insulation plate may cause leakage
current.
Clean the insulation plate as described in “Cleaning the Insulation Plate” on
page 207.

Table of Contents

Other manuals for Agilent Technologies 7100

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7100 and is the answer not in the manual?

Agilent Technologies 7100 Specifications

General IconGeneral
BrandAgilent Technologies
Model7100
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals