222 Agilent 7100 Capillary Electrophoresis System User Manual
8Maintenance
Changing the Air Inlet Filter
Changing the Air Inlet Filter
Objective
The following procedure is used to replace the air inlet filter. The installed air
pump uses ambient air (through this filter) to create pressure inside the
electrolyte bottle. The compressed air is needed to flush a capillary and to
inject a sample.
Frequency
Air filter replacement should be considered if the environment is very dusty
or if the system cannot generate enough pressure. We suggest that you change
the air filter at least every three months.
Accessing the Air Filter
The air filter is located inside the electrolyte and waste bottle compartment,
see Figure 71 on page 223. Therefore, it is very easy to replace the air filter
frequently.
If the instrument cannot build-up either pressure or a vacuum, check the
following:
proper seal of the pressure bottle,
proper seal of the vacuum bottle, and
loose fittings on the bottles
If none of these reasons apply, call Agilent Technologies.

Table of Contents

Other manuals for Agilent Technologies 7100

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7100 and is the answer not in the manual?

Agilent Technologies 7100 Specifications

General IconGeneral
BrandAgilent Technologies
Model7100
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals