EasyManuals Logo
Home>Agilent Technologies>Laboratory Equipment>7100

Agilent Technologies 7100 User Manual

Agilent Technologies 7100
266 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #205 background imageLoading...
Page #205 background image
Agilent 7100 Capillary Electrophoresis System User Manual 205
Maintenance 8
Cleaning the Electrodes, Pre-punchers and Insulation Plate
Cleaning the Electrodes
1 Flush the electrodes with water. Watch for salt crystals. All salt crystals
must be removed.
Figure 64 Flushing electrodes
2 Flush the electrodes with isopropanol.
3 Use compressed air to dry the electrodes and remove all traces of liquid.
CAUTION
Increased leakage current and arcing due to wet electrodes
Increased leakage current and arcing might appear during analysis if electrodes are not
completely dry.
To avoid contamination, the inside of the electrodes must be completely dry before
reinstallation.
NOTE
If there is still an accumulation of dirt in the upper funnel or inside the electrode start the
cleaning procedure again. You may place it in a beaker of water or isopropanol and treat this
in an ultrasonic bath for 5 minutes followed by steps 2 and 3 in this section. If this fails,
replace the electrode.

Table of Contents

Other manuals for Agilent Technologies 7100

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7100 and is the answer not in the manual?

Agilent Technologies 7100 Specifications

General IconGeneral
BrandAgilent Technologies
Model7100
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals