Agilent 7100 Capillary Electrophoresis System User Manual 207
Maintenance 8
Cleaning the Electrodes, Pre-punchers and Insulation Plate
Cleaning the Insulation Plate
Salt crystals, films of substances such as buffer or gel that allow moisture will
lead to increased leakage current or even arcing. The insulation plate must be
free of such contamination.
1 Moisten a tissue with water and use it to clean the insulation plate.
2 Moisten a tissue with isopropanol and use it to clean the insulation plate.
3 Dry the insulation plate with a tissue and compressed air.
NOTE
If there is an accumulation of dirt in the pre-puncher funnel, put the pre-punchers and
connecting tubing in a beaker of water and treat it in an ultrasonic bath for five minutes
followed by steps 2 and 3 in this section.
NOTE
Residual liquid in connecting tubes may be transported to the air valve blocks and result in
damage. Such liquid plugs may also negatively influence injection reproducibility.

Table of Contents

Other manuals for Agilent Technologies 7100

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7100 and is the answer not in the manual?

Agilent Technologies 7100 Specifications

General IconGeneral
BrandAgilent Technologies
Model7100
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals