254 Agilent 7100 Capillary Electrophoresis System User Manual
10 Appendix
Setting Up a Test Sample Run
Preconditioning
CE Conditioning
Flush inlet 1, outlet 8 for 5 minutes
Wait 5 minutes
Flush inlet 2, outlet 8 for 5 minutes
Flush inlet 6, outlet 8 for 10 minutes
CE Injection
Inject by pressure, 50 mbar, 4 seconds
Timetable
Change voltage 0.2 minutes, 30 kV
Postconditioning
No postconditioning used