68 Agilent 7100 Capillary Electrophoresis System User Manual
4 LAN Configuration
Initialization Mode Selection
Since the default IP address is a so-called local address, it will not be routed by
any network device. Thus, the PC and the CE instrument must be in the same
subnet.
The user may open a Telnet session using the default IP address and change
the parameters stored in the non-volatile memory of the CE instrument. He
may then close the session, select the initialization mode "Using Stored",
power-on again and establish the TCP/IP connection using the new
parameters.
When the 7100 CE instrument is wired to the PC directly (for example using a
cross-over cable or a local hub), that is separate from the local area network,
the user may simply keep the default parameters to establish the TCP/IP
connection.
Table 19 Using Default parameters
IP address: 192.168.254.11
Subnet mask: 255.255.255.0
Default gateway Not specified
NOTE
In the "Using Default" mode, the parameters stored in the memory of the 7100 Capillary
Electrophoresis instrument are not cleared automatically. If not changed by the user, they
are still available when you switch back to the mode "Using Stored".

Table of Contents

Other manuals for Agilent Technologies 7100

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7100 and is the answer not in the manual?

Agilent Technologies 7100 Specifications

General IconGeneral
BrandAgilent Technologies
Model7100
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals