Agilent 7100 Capillary Electrophoresis System User Manual 89
Using the Agilent 7100 Capillary Electrophoresis System 5
Working with the CE Diagram
Direct Control of Instrument Components
The actions are performed immediately and the changes are reflected in the
CE diagram. Such actions include, for example:
Switching the lamp on.
Controlling the sample tray.
Lowering the lift stations to remove the capillary cassette.
Releasing the pressure to change the replenishment bottles.
Working with these items does not change the method currently loaded.
Set Error Method
Set Error Method enables you to store the current parameters. These parameters
are applied to the instrument when an error occurs. It ensures that the
instrument shuts down in a controlled manner if ChemStation control is
discontinued for any reason.
Identify Device
The command Identify Device causes the LED on the front cover to blink
alternating in green and yellow for 4 seconds.