152
Agilent 7820 GC Service
PP in the Maintenance manual.)
 Set the oven, inlet, and detector to 250 °C and bake out
for at 15 min. (See the procedure for the SS or PP in
the Maintenance manual.)
Be sure to configure the column.
1 If it is not already installed, install the phosphorus filter.
Create or load a method with the parameter values listed in the
table below.
FPD Checkout Conditions (P)
HP-5, 30 m x 0.32 mm x 0.25 µm
(19091J-413)