Agilent 7820 GC Service
Hazardous voltages are present in the
mainframe when the GC power cord is connected.
Avoid a potentially dangerous shock hazard by
disconnecting the power cord before removing
any GC panels.
Components can be damaged by static electricity:
be sure to wear an ESD strap grounded to the GC
chassis while performing this procedure.
Inlets, detectors, and the oven are insulated
with fibrous materials which may cause
irritation to skin, eyes, and/or mucous
membranes. Always wear gloves when working
with the insulation. Additionally, if the
insulation is flaky/crumbly, wear protective
eyewear and a suitable breathing mask and/or
respirator.
Heater replacement requires removal of the inlet. Refer to
Replacing the Purged Packed inlet (31) for removal
instructions.