D AssayMAP Bravo Platform
Requirements for partial-head liquid handling
222 Bravo Platform User Guide
Although the VWorks software considers the geometry of the labware at the target deck
location during the task, the software is unaware of the labware height on the adjacent
deck location where the head will be offset.
Requirements for mounting a partial head of 250-µL pipette tips
CAUTION
A collision will result if the Bravo 96AM Head is offset from the tip box location when
pressing on pipette tips. When picking up pipette tips from a tip box, ensure that the
Bravo 96AM Head is centered over the tip box.
CAUTION
To ensure consistent seating of pipette tips on the probes, use the 96AM Cartridge &
Tip Seating Station to seat the pipette tips on the probes.
The Bravo 96AM Head must be centered over the tip box when picking up pipette tips to
be transfered to the 96AM Cartridge & Tip Seating Station.
Figure
96AM Cartridge & Tip Seating Station
Related information
For more information about… See…
Bravo 96AM Head “Bravo 96AM Head overview” on page 200
Using and storing the Bravo 96AM
Head
“Installing and using the Bravo 96AM Head”
on page 208

Table of Contents

Other manuals for Agilent Technologies Bravo Platform

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies Bravo Platform and is the answer not in the manual?

Agilent Technologies Bravo Platform Specifications

General IconGeneral
BrandAgilent Technologies
ModelBravo Platform
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals