A Accessories and platepads
Installing and setting up an Orbital Shaking Station
106
Bravo Platform User Guide
Installing and setting up an Orbital Shaking Station
Description
The Orbital Shaking Station is a microplate shaker that can be installed at any deck
location on top of a short adapter pad instead of a platepad. The Orbital Shaking Station
has the option of using its own control module, or an Inheco Single TEC controller (STC)
or Multi TEC controller (MTC). For a full description of the Orbital Shaking Station, see
the manufacturer’s documentation.
Figure
Orbital Shaking Station with its own control module and an adapter pad
Setup workflow
Step Procedure See…
1 Install the Orbital Shaking Station. “Installing the Orbital Shaking
Station” on page 107
2 Configure the station in Diagnostics. “Configuring the Orbital Shaking
Station in Bravo Diagnostics” on
page 110
3 Test the station operation. “Testing the Orbital Shaking
Station” on page 111
4 Edit the teachpoint using the pipettor
teach plate. Be sure to consider the 10-
mm thickness of the teach plate when
you edit the teachpoint.
“Adjusting the teachpoint using
a teach plate” on page 113

Table of Contents

Other manuals for Agilent Technologies Bravo Platform

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies Bravo Platform and is the answer not in the manual?

Agilent Technologies Bravo Platform Specifications

General IconGeneral
BrandAgilent Technologies
ModelBravo Platform
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals