D AssayMAP Bravo Platform
Bravo 96AM Head overview
200 Bravo Platform User Guide
Bravo 96AM Head overview
This topic describes:
• “Bravo 96AM Head components” on page 200
• “Liquid-handling modes” on page 202
Bravo 96AM Head components
The Bravo 96AM Head components include:
• 96 channels for liquid handling. Each channel consists of a syringe with a probe on
the end. A syringe consists of a glass barrel with a plunger and a seal. The syringes
can employ a combination of direct-liquid-flow and air-displacement techniques for
liquid-handling tasks.
The following figure shows a syringe and a cross-sectional close-up view of the
plunger in the probe.
Bravo 96AM Head • “Bravo 96AM Head overview” on
page 200
• “Installing and using the Bravo 96AM
Head” on page 208
• “Requirements for partial-head liquid
handling” on page 219
AssayMAP Bravo cartridges “AssayMAP cartridge overview” on
page 205
Cartridge & Tip Seating Station “About seating the cartridges or pipette tips
on the probes” on page 206
For more information about… See…