B Autofilling accessories
Setting up a Weigh Station
182 Bravo Platform User Guide
Setting up a Weigh Station
About this topic
This topic provides calibration instructions for the Weigh Station.
For a description of the Weigh Station and installation instructions, see the Pump
Module User Guide.
Calibrating a Weigh Station
Before you begin
Make sure the Pump Module is set up and connected to the Weigh Station. See the
Pump Module User Guide.
Verify that the Autofill Station accessory is configured in Bravo Diagnostics. See
“Configuring an autofilling station for the Bravo Platform” on page 177.
Procedure
Before the Weigh Station is ready for use in a protocol, you must calibrate the empty
and full settings. When you calibrate the Weigh Station, you must use the reservoir type
that will be weighed.
To calibrate a Weight Station:
1 In Diagnostics, click the Profiles tab and make sure the correct profile is initialized.
2 Click the Configuration tab.
3 In the Accessory Configuration area, highlight Autofill Station, and then click
Diagnose accessory.
Note: The Autofill Station is represented by a Pump Module graphic in the
software.
00268
Bravo
weigh pad
Crosshairs for setting
teachpoint

Table of Contents

Other manuals for Agilent Technologies Bravo Platform

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies Bravo Platform and is the answer not in the manual?

Agilent Technologies Bravo Platform Specifications

General IconGeneral
BrandAgilent Technologies
ModelBravo Platform
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals