5 Using diagnostics software
Using the Move and Approach commands
Bravo Platform User Guide
Using the Move and Approach commands
About this topic
The Bravo Move and Approach commands let you quickly move the liquid-handling
head to any of nine locations on the deck.
Typically, these commands are used when editing teachpoints. This topic describes
how to move to and approach locations on the Bravo deck using these commands.
WARNING
To avoid potential injury and damage to the device, only personnel trained in how to
teach the Bravo Platform should use the Jog/Teach tab controls.
WARNING
The red Stop motors button in Bravo Diagnostics does not perform an immediate
stop. The Bravo head can continue to move in the same direction at the same speed
after you click the button. To perform an emergency stop, press the red button on the
emergency-stop pendant.
Commands defined
Move. Instructs the liquid-handling head to move to the teachpoint for the selected
location.
Approach. Instructs the liquid-handling head to move to the teachpoint for the
selected location, with a vertical offset equal to the value specified in the Approach
height above teachpoint box.
Note: The Approach height parameter in the profile is the height above the
labware when pipetting.
Before you start
Make sure no labware is on the Bravo deck and that the liquid-handling head has been
taught with the same type of tips used in the profile.
WARNING
Before you command the Bravo head to move, make sure that everyone is clear of
the Bravo Platform.
CAUTION
Before using the Move or Approach command, make sure the path is clear of any
obstacles that you have not defined in the profile. The move-to-position commands
present a higher damage risk to the liquid-handling head, because the head can
move in large increments.
Procedure
To use the Move command:
1 Open Diagnostics, and then click the Jog/Teach tab.
2 Select a deck location from the Location box.

Table of Contents

Other manuals for Agilent Technologies Bravo Platform

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies Bravo Platform and is the answer not in the manual?

Agilent Technologies Bravo Platform Specifications

General IconGeneral
BrandAgilent Technologies
ModelBravo Platform
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals