AM Dispense task parameters
Bravo Platform User Guide
Dispense flow rate (0–500
µL/s)
The numerical value or the JavaScript variable
that will override the dispense velocity setting in
the liquid class.
A JavaScript variable enables the value to be
assigned later. For example, using a VWorks form,
an operator could easily change the flow rate for
an aspirate step in increments from as low as
1 µL/min up to 2000 µL/min or more using the
same liquid class.
If the task is included in a VWorks macro, a
JavaScript variable enables you to change the
value for the task at the macro level.
IMPORTANT The software requires that the
flow rate value be in microliters per second (µL/s)
at run time. If you want an operator to enter the
value in microliters per minute (µL/min), you can
use scripting to convert the values for the
software to use.
Parameter Description

Table of Contents

Other manuals for Agilent Technologies Bravo Platform

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies Bravo Platform and is the answer not in the manual?

Agilent Technologies Bravo Platform Specifications

General IconGeneral
BrandAgilent Technologies
ModelBravo Platform
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals